Reference Materials from: VSL

Reset

                           


Reference Materials from: VSL

Producer   RM Code   Name    Item   Trust Updated  
  Properties:
Benzene (8CI, 9CI) (CAS 71-43-2) ≥ 1 nmol/mol
Benzene (8CI, 9CI) (CAS 71-43-2) ≤ 1 µmol/mol
Toluene (CAS 108-88-3) ≥ 1 nmol/mol
Toluene (CAS 108-88-3) ≤ 1 µmol/mol
Ethylbenzene (CAS 100-41-4) ≥ 1 nmol/mol
Ethylbenzene (CAS 100-41-4) ≤ 1 µmol/mol
o-Xylene (CAS 95-47-6) ≥ 1 nmol/mol
o-Xylene (CAS 95-47-6) ≤ 1 µmol/mol
m-Xylene (CAS 108-38-3) ≥ 1 nmol/mol
m-Xylene (CAS 108-38-3) ≤ 1 µmol/mol
p-Xylene (CAS 106-42-3) ≥ 1 nmol/mol
p-Xylene (CAS 106-42-3) ≤ 1 µmol/mol

Remarks:
Primary Reference Materials containing BTEX in nitrogeb are available in various compositions. Nominal fraction range:2 ppb to 1 ppm (mol/mol). Within the listed range any composition can be supplied. Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3

Intended Use:
This Primary Reference Materials is intended to be used for calibrating instruments and evaluating analytical methods. E.g.: Determination of all kinds of emissions in the environment, such as pollutants and greenhouse gases; automotive test gases which are prescribed in OIML recommendation R 99 “Instruments for measuring vehicle exhaust emissions”; etc. etc.

CRM Certificate:  Example PRM.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/121000

  Properties:
Krypton * (Kr) (CAS 7439-90-9) 10 µmol/mol
Argon (8CI, 9CI) * (Ar) (CAS 7440-37-1) 10 µmol/mol
Neon * (Ne) (CAS 7440-01-9) 10 µmol/mol

Remarks:
Primary Reference Materials containing Argon, Neon and Krypton in Helium with nominal fraction of 10 ppm (mol/mol) of each component. Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3. Other compositions, including other components, are available on request.

Intended Use:
This Primary Reference Materials is intended to be used for calibrating instruments and evaluating analytical methods. E.g.: Determination of all kinds of emissions in the environment, such as pollutants and greenhouse gases; automotive test gases which are prescribed in OIML recommendation R 99 “Instruments for measuring vehicle exhaust emissions”; etc. etc.

CRM Certificate:  Example PRM.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/121050

  Properties:
Carbon oxide (CO2) (CAS 124-38-9) 10 % mol/mol
Oxygen * (O2) (CAS 7782-44-7) 90 % mol/mol

Remarks:
Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3 depending on the concentration.

Intended Use:
The listed Primary Reference Materials (PRMs) are typical examples of standard gas mixtures used in blood gas determination, respiratory and anesthesia monitoring. Other compositions, including other components, are available on request.

CRM Certificate:  Example PRM.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/127200

  Properties:
Carbon oxide - CO2 - (CAS 124-38-9) 5 % mol/mol
Oxygen - O2 - (CAS 7782-44-7) 95 % mol/mol

Remarks:
Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3 depending on the concentration.

Intended Use:
The listed Primary Reference Materials (PRMs) are typical examples of standard gas mixtures used in blood gas determination, respiratory and anesthesia monitoring. Other compositions, including other components, are available on request.

CRM Certificate:  Example PRM.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/127250

  Properties:
Nitrous Oxide * (N2O) (CAS 10024-97-2) 50 %mol/mol
Oxygen * (O2) (CAS 7782-44-7) 20 %mol/mol

Remarks:
This Primary Reference Materials contains nitrous oxide and oxygen. Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3. Other compositions are available on request.

Intended Use:
The Primary Standard Gas Mixtureis a typical example of standard gas mixtures used in blood gas determination, respiratory and anesthesia monitoring.

CRM Certificate:  Example PRM.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/129100

  Properties:
Helium * (He) (CAS 7440-59-7) 10 µmol/mol

Remarks:
The listed PRM is a typical example of standard gas mixtures used in leak detection.. Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3. Other compositions, including other components, are available on request.

Intended Use:
The composition listed are on leakage measurement level. Some of these gases are used as tracers in dispersion modelling

CRM Certificate:  not yet available

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/217250

  Properties:
Ammonia * (NH3) (CAS 7664-41-7) ≥ 10 µmol/mol
Ammonia * (NH3) (CAS 7664-41-7) ≤ 5000 µmol/mol

Remarks:
Certified reference materials containing ammonia in nitrogen are available in a range of composition. Nominal fraction range: 30 ppm to 300 ppm (mol/mol). Within the listed range any composition can be supplied. The Certified Reference Material (CRM) will be prepared according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures" but the certified composition will be given on the certificate according to ISO 6143:2001 "Gas analysis - Comparison methods for determining and checking the composition of calibration gas mixtures". Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3

Intended Use:
This Primary Reference Materials is intended to be used for calibrating instruments and evaluating analytical methods. E.g.: Determination of all kinds of emissions in the environment, such as pollutants and greenhouse gases; automotive test gases which are prescribed in OIML recommendation R 99 “Instruments for measuring vehicle exhaust emissions”; etc. etc.

CRM Certificate:  Example CRM NH3.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/217300

  Properties:
Carbon oxide (CO2) (CAS 124-38-9) ≥ 1 % mol/mol
Carbon oxide (CO2) (CAS 124-38-9) ≤ 18 % mol/mol
Carbon monooxide (CO) (CAS 630-08-0) ≥ 0.1 % mol/mol
Carbon monooxide (CO) (CAS 630-08-0) ≤ 9 % mol/mol
Propane (C3H8) (CAS 74-98-6) ≥ 0.005 % mol/mol
Propane (C3H8) (CAS 74-98-6) ≤ 0.02 % mol/mol
Oxygen (O2) (CAS 7782-44-7) ≥ 0.1 % mol/mol
Oxygen (O2) (CAS 7782-44-7) ≤ 22 % mol/mol

Remarks:
Primary Reference Materials containing typical Automotive Gas are available in various compositions. Within the listed range any composition can be supplied. Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3

Intended Use:
This Primary Reference Materials is intended to be used for calibrating instruments and evaluating analytical methods. E.g.: Determination of all kinds of emissions in the environment, such as pollutants and greenhouse gases; automotive test gases which are prescribed in OIML recommendation R 99 “Instruments for measuring vehicle exhaust emissions”; etc. etc.

CRM Certificate:  Example PRM.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/279550

  Properties:
Nitrogen dioxide - NO2 - (CAS 10102-44-0) ≥ 10 µmol/mol
Nitrogen dioxide - NO2 - (CAS 10102-44-0) ≤ 1000 µmol/mol

Remarks:
Primary Reference Materials containing nitrogen dioxide in nitrogen are available in various compositions. Nominal fraction range: 10 ppm to 1000 ppm (mol/mol). Within the listed range any composition can be supplied. Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3

Intended Use:
This Primary Reference Materials is intended to be used for calibrating instruments and evaluating analytical methods. E.g.: Determination of all kinds of emissions in the environment, such as pollutants and greenhouse gases; automotive test gases which are prescribed in OIML recommendation R 99 “Instruments for measuring vehicle exhaust emissions”; etc. etc.

CRM Certificate:  Example PRM in N2.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/308200

  Properties:
Nitrite NO2 - * nitrite ions (CAS 14797-65-0) ≥ 10 µmol/mol
Nitrite NO2 - * nitrite ions (CAS 14797-65-0) ≤ 1000 µmol/mol

Remarks:
Primary Reference Materials containing nitrogen dioxide in synthetic air are available in various compositions. Nominal fraction range: 10 ppm to 1000 ppm (mol/mol). Within the listed range any composition can be supplied. Each mixture is prepared individually according to ISO 6142-1:2015 "Gas analysis - Preparation of calibration gas mixtures - Part 1: Gravimetric method for Class I mixtures". After preparation the composition is verified against the National Primary Standard Gas Mixtures of the Netherlands. Cylinder Pressure 11,5 MPa Cylinder Valve Outlet DIN 477 No 1 Approximate deliverable volume 0,6 m3

Intended Use:
This Primary Reference Materials is intended to be used for calibrating instruments and evaluating analytical methods. E.g.: Determination of all kinds of emissions in the environment, such as pollutants and greenhouse gases; automotive test gases which are prescribed in OIML recommendation R 99 “Instruments for measuring vehicle exhaust emissions”; etc. etc.

CRM Certificate:  Example PRM in Synthetic air.pdf

Producer: (click for contact & purchase details)
 VSL (Netherlands)

This material is produced under a valid ISO 17034 accreditation | This material is linked to statutory activities as a national metrology institute (NMI) / a designated institute (DI)

 

Public link to this factsheet: www.comar.bam.de/rm/308450